Beamline hardware issues

Search and lock procedure times out before procedure is completed

  1. Try Search and lock procedure again. The procedure is timed so make sure you are doing the procedure in good time.

  2. Check all search buttons are being pressed in correct order.

  3. Check that you are not setting of light gates (on hutch door) before starting procedure.

  4. Call BL staff/EHCs.

Expand
titleSearch and lock procedure times out before procedure is completed 
  1. Try Search and lock procedure again. The procedure is timed so make sure you are doing the procedure in good time.

  2. Check all search buttons are being pressed in correct order.

  3. Check that you are not setting of light gates (on hutch door) before starting procedure.

  4. Call BL staff/EHCs.

Expand
titleSearch procedure wont start 
  1. Check all doors are correctly closed

Expand
titleThe shutter is not opening
  1. Has it been more than 3 minutes? 
    If you have not been taking beam into the hutch recently, GDA will wait 3 minutes to open the shutter.

image-20240628-143911.png
  1. Check the gas levels (User information screen).

    1. The PCO gas (Argon Technical Grade) should flow at 1.7 SCCM

      1. The KB Gas (BIP Helium) should flow at 3.5 SCCM

      2. Check regulators on the bottles themselves.

    2. If gas has run out call EHC’s to replace (8787) – let them know which gas and which grade (Argon – Technical, Helium – BIP).

  2. Check shutter interlocks

    1. On User information screen check interlock icons - all should be bright green after opening the shutter.

    2. If Guardlines Status is not bright green - Search and lock the hutch again.

    3. For anything else contact BL Staff (Through EHCs if out of hours).

  3. Otherwise Call BL STAFF

Expand
titleGTS X,Y,Z aren't moving
  1. Call BL Staff

Expand
titleDiffraction detector icon is red

Expand
titleImaging detector icon is red

Scan issues

Expand
titleIf you cannot see your sample in the imaging FOV

Check the shutter is open. 

  1. If shutter fails to open check page for EH shutter issues.

    image-20240628-144210.png

  2. Check the beam selector is set to correct beam.

    1. The top drop-down field (user selection) and the bottom field (read back/actual beam) can become out of sync. (eg. Top says imaging but readback says diffraction).

    2. In this case – set the drop down menu to the same as the read back value and then again use the drop down menu to select the beam of choice.

    image-20240628-144230.png

  3. Check that the camera stream is on. 

    1. Turn on camera stream (1) 
      The camera button should go light green (shown below). If it does not go green

    2. Initiate a new camera (2) 
      This should open a new window.

    3. Using stream drop down select EPICS PV Access and press connect. Once you click connect, you should see a live stream.

    4. If the imaging camera or pilatus does not initiate/turn on or icon goes red Call BL Staff.

      image-20240628-144323.png

  4. Check that the exposure time is set correctly. 

    1. If the exposure is too high, the detector will max out (65K pixel count for imaging). This is not good for either detector so stop ASAP. If it is too low then the image will appear black.

  5. Check your sample in the correct position. 

    1. Rotate sample 180 degrees using GTS theta – the sample should appear in the view

    2. Check base-X is set to GTS position

    3. Check the sample height is correct. Use the laser as a guide.

Expand
titleDiffraction scan failing