Search procedure wont start 

  1. Check all EH doors are closed properly

Search and lock procedure times out before procedure is completed

  1. Try Search and lock procedure again. The procedure is timed so make sure you are doing the procedure in good time.

  2. Check all search buttons are being pressed in correct order.

  3. Check that you are not setting of light gates (on hutch door) before starting procedure.

  4. Call BL staff/EHCs.

The Shutter isn’t opening

  1. Has it been more than 3 minutes? 
    If you have not been taking beam into the hutch recently, GDA will wait 3 minutes to open the shutter.

  2. Check the gas levels (User information screen).

    1. The PCO gas (Argon Technical Grade) should flow at 1.7 SCCM

      1. The KB Gas (BIP Helium) should flow at 3.5 SCCM

      2. Check regulators on the bottles themselves.

    2. If gas has run out call EHC’s to replace (8787) – let them know which gas and which grade (Argon – Technical, Helium – BIP).

  3. Check shutter interlocks

    1. On User information screen check interlock icons - all should be bright green after opening the shutter.

    2. If Guardlines Status is not bright green - Search and lock the hutch again.

    3. For anything else contact BL Staff (Through EHCs if out of hours).

  4. Otherwise Call BL STAFF

image-20240628-143911.pngImage Removed

GTS X,Y,Z aren't moving

  1. Call BL Staff

Diffraction detector icon is red

Base-X isn’t moving

...

Issues with troubleshooting procedure:

Child pages (Children Display)
depth1
allChildrentrue
style
sortAndReverse
first0

Issues to call BL staff directly:

  1. GTS motors not moving/responding

  2. Base-X not moving/responding