PLEASE USE THIS GUIDE AND FOLLOW THE ACTIONS ONLY IF YOU FEEL CONFIDENT TO DO SO. Phone Numbers Emergency - 2222 EHCs - 8787 Control room 8899 IF NOT PLEASE CALL THE BL STAFF OR IF OUT OF HOURS CALL THE EHC’S WORKING HOURS: 0900-1700 |
Beamline hardware issues
Expand | ||
---|---|---|
| ||
Search and lock procedure times out before procedure is completed
Search procedure wont start
|
...
title | The shutter is not opening |
---|
Has it been more than 3 minutes?
If you have not been taking beam into the hutch recently, GDA will wait 3 minutes to open the shutter.Check User information screen (on TV)
...
Check the gas levels (User information screen).
The PCO gas (Argon Technical Grade) should flow at 1.7 SCCM
The KB Gas (BIP Helium) should flow at 3.5 SCCM
Check regulators on the bottles themselves.
If gas has run out call EHC’s to replace (8787) – let them know which gas and which grade (Argon – Technical, Helium – BIP).
Check shutter interlocks
On User information screen check interlock icons - all should be bright green after opening the shutter.
If Guardlines Status is not bright green - Search and lock the hutch again.
For anything else contact BL Staff (Through EHCs if out of hours).
Otherwise Call BL STAFF
Scan issues
Expand | ||
---|---|---|
| ||
Check the shutter is open.
|