  1. At the start of the day perform a high flash and run up the emission.
  2. Open the beam valve and check there is a beam on the large phosphor screen.
  3. Ensure that ‘Ronchigram’ is selected on TEMCentre and that scanning is stopped on digital micrograph.
  4. Check objective lens is set to standard defocus.
  5. Bring sample to eucentric height (infinite magnification).
  6. Set camera length to 20cm.
  7. Manually correct for coma and two-fold astigmatism in the Ronchigram on the focus phosphor screen or camera.
  8. Centre the Ronchigram using the projector lens deflectors.
  9. If necessary run the corrector (CEOS for E01, COSMO for E02)
  10. Insert the required condenser lens aperture and centre to the flat area of the ronchigram.
  11. Select the required camera length.
  12. Using the projector lens deflectors centre the bright field disc to the centre of the detectors.
  13. Deselect ‘Ronchigram’ in TEMCentre.
  14. In TEMCentre, select the required detectors for imaging.  