In the current DAWN release (1.6) a few steps must be taken to correctly set up the Jython interpreter to run a calibration. These steps will likely be unnecessary in future releases.
Add following Jars from DAWN installation plugin folder to Jython path (Window - Preferences - PyDev - Interpreter - Jython Interpreter - New Jar):
uk.ac.diamond.scisoft.diffraction.powder_version.jar
uk.ac.diamond.org.jscience4_version.jar
com.throughtworks.xstream_version.jar
Once these Jars are included, the following script may be used to calibrate an image.
import scisoftpy as dnp from uk.ac.diamond.scisoft.diffraction.powder import PowderCalibration as Cal from uk.ac.diamond.scisoft.analysis.crystallography import CalibrationFactory as Fac from uk.ac.diamond.scisoft.analysis.diffraction.powder import PixelSplittingIntegration as PSI #pixel size in mm, number of rings to use for calibration (from centre) pixelSize = 0.2 numberOfRings = 6 #Load data (and plot in "Plot 1") data = dnp.io.load(r"/Path/To/Calibration/File.tiff")["image_01"] dnp.plot.image(data) #Get standard, if using a new standard, configure in DAWN first #Use name of standard in DAWN list std = Fac.getCalibrationStandards() ceo2 = std.getCalibrationPeakMap("CeO2") #Calibrate, get diffraction metadata object out - required for integration routines. out = Cal.calibrateSingleImage(data._jdataset(),pixelSize,ceo2.getHKLs(),numberOfRings) print "Residual from fit: " + str(out.getResidual()) meta = Cal.createMetadataFromOutput(out, 0, data.shape, pixelSize) #Integrate integrator = PSI(meta, 2000) result = integrator.integrate(data._jdataset()) dnp.plot.line(result[0], result[1])