E01 is equipped with a Gatan Quantum GIF for electron energy loss spectroscopy.


  1. Set camera length to 4cm.
  2. Select 5mm entrance aperture on EELS spectrometer.
  3. Select dispersion value required.
  4. Select user mode.
  5. Set exposure time to something small.
  6. Adjust PL to maximise intensity of zero loss peak.
  7. Adjust focus x, y and Sx , Sy of zero loss peak.
  8. For data collection use SNR mode or HR mode.
  9. Once spectrum obtained highlight area around zero-loss peak, select SI menu and align SI by peak.

 